What are the specifications of electrical cables from the main power source to the ESWL room? Is there any recommendation for the power cord specifications cross sectional diameter, etc?

For the ESWL Room we recommend using a flexible PVC insulated multi-strand copper wire of the size 10 AWG each for phase, neutral and ground wire. When using a 3 core cable instead of single conductors, we recommend a cable having 3 conductors of size 2.5mm diameter excluding the PVC shield. This calculation is suitable for TMC-Compact with in-built c-arm. However if you are using any other make c-arm along with TMC-SA, we request you to contact the c-arm manufacturer for details.