This procedure may not be applicable with all the models of ESWL equipment manufactured by TMC as some equipment come with an inbuilt imaging system. This procedure is suitable for Lithotripters that do not have an inbuilt imaging system and the user wants to use a imaging system ( a C-Arm image intensifier ) that is available in most hospitals. Alignment of the C-arm to the ESWL is achieved using alignment sights on the ESWL and a “cross” to be positioned on the image intensifier. The lower sight is already mounted on the ESWL, it will appear as a “crossbar” on the X-Ray image. The upper sight will appear as a “pointer” on the X-Ray image. The middle sight, pointing to the focal zone, is positioned on the reflector open face and will appear as an “arrow” on the X-Ray image. Place the “cross” at the center of the C-arm image intensifier face. The “cross” now defines the center of the display monitor.
The C-arm/ESWL alignment is done using five movable axes on the C-arm: Wig-Wag axis; boom axis; Pitch axis; Roll axis; and the Up-Down axis. The alignment procedure is as follows:
a) Initial position of C-arm axes is as follows:
Boom - Set at Midway of axis. Wig-Wag - Set at 0 degrees. Roll - Set at 0 degrees. Pitch - Set at 0 degrees. Up-Down - Left as is.
b) Turn the C-arm on. c) Move Boom and Wig-Wag axes to position the “arrow” tip (focal sight) to the center of the display monitor. d) Using the display monitor controls, align the vertical axis of the monitor with the orientation of the patient’s spine such that the patient’s head is at the top, and the patient’s left side is on the right side of the screen. e) Verify that the arrow is still at the center of the display monitor. If not, perform step c, again. f) Lock Boom and Wig-Wag axes in place. g) Move the Roll and Pitch alternately until the pointer (upper sight), the crossbar (lower sight) and the arrow (focal sight) are all set. h) Lock Roll and Pitch in place. i) Move the roll to about a 30 degree angle. j) Using the C-arm controls adjust the height of the C-arm so that the “cross” on the display is at the tip of the arrow. k) Move Roll back to 0 degrees and check to make sure the crossbar, the pointer, and the arrow are set as before, and that the “cross” on the display is still where the crossbar and the pointer meet. l) The C-arm and the ESWL are now aligned. m) Remove the upper and focal sights from the ESWL and place them aside gently. You are now ready to go... |
|
|
|
Search Knowledge Base Articles
What is the procedure of alignment of c arm with ESWL machine (lithotripter).
Did you find this article useful?
Related Articles
-
Do we have to change the distilled water in ESWL machine every case and if there is a filter in the machine for using the distilled water for longer time ?
The distilled water does not require change after every treatment and can be used for 10 to 12 sessi... -
Does your ESWL machine, lithotripter has the ecg and respiratory triggering options?
As a standard feature our ESWL machine comes with a simulation based triggering option. We also prov... -
I want the service manual for lithotripter?
As per our standard company policy we do not provide the customers with a service manual for any pro... -
I want the spare parts list for lithotripter?
The spares part list is not provided to any users and is only available with authorized service cent... -
Is it necessary to give anesthesia to the patient during lithotripsy treatment on TMC manufactured ESWL machines ?
No anaesthesia is required during a Lithotripsy session in any of TMC ESWL machines. However, in ...