The ‘IECP’ machine is warranted for a period of 12 months from the date of supply against any ma...
-
During the daily use of the IECP machine, there are some things which will be required on regular basis, and are termed as consumables. These would include • Set of Cuffs – a set of cuffs co...
-
IECP is an OPD procedure and does not require much of the place for therapy. The room size required for IECP system is 8 Feet X 10 Feet. Other requirements of IECP equipment include : A dedicated IECP room of size of ...
-
For the IECP Room we recommend to use a flexible PVC insulated multi-strand copper wire of the size 10 AWG e...
-
The ‘IECP’ machine manufactured by TMC can perform limited self servicing on its own periodicall...
-
Power rating of IECP: SL - 200 Voltage 415VAC (~10%), Three Phase / Four wire Frequency 50Hz Load 2.25 KW Normal Current consumption per phase 6 Amp Maximum Current consumption per phase 7.5 Amp...
-
If you are going to or have already purchased the IECP machine from TMC, then the installation of IECP machine will be done at your site without any extra charge by qualified TMC engineer(s). Some of the requirements for installation of IECP ma...
-
If you are going to or have already purchased an IECP machine from TMC, the training on use of IECP equipment will be provided by TMC at no extra charge. TMC’s engineer / technician will be deputed to your centre for a...